Dry Cabinet Autoclave
1.This series of dry cabinet autoclave is used for the high temperature sterilization and depyrogenation to the stable material and items after cleaning by successive circulation of high temperature air.
2.This series of autoclave adopts horizontal rectangular structure.
3.304 or 316L stainless steel chamber, covered by whole stainless steel plate, and equipped with standard GMP verification interface.
4.Industrial control system PLC + HMI has a standard “user management, data storage, process management” function. The sterilization process can be traced back, with independent data logging device and industrial control system to ensure data accuracy.
5.Dry Cabinet Autoclave Principle:
High efficiency filter is installed on both air inlet port and outlet port side of the dry autoclave, filtration efficiency is 99.99%. Fresh air port and wet discharge port is equipped with multi-wing centrifugal fan. It can inlet air and discharge heat automatically. Flat flow hot air circulation device fan keeps the temperature even. Stainless steel convection wall on two sides of the chamber could be disassembled, cleaned and operated conveniently.
6.Scope of Dry Cabinet Autoclave Application:
This dry autoclave is widely used for the high-temperature sterilization operation of injection ampule, vials, oral liquid bottles and other medical glass bottles and stainless steel utensils in pharmaceutical factories.
The DMH series autoclave is dry cabinet autoclave, which is mainly used is used for the high temperature sterilization and depyrogenation to the stable material and items after cleaning by successive circulation of high temperature air.