Correct Operation Specification Of Hydrogen Peroxide Sterilizer

Correct Operation Specification Of Hydrogen Peroxide Sterilizer:

As a medical sterilizer, the tuttnauer  hydrogen peroxide sterilizer can quickly sterilize medical equipment through reasonable application.

Although the sterilization effect of  the tuttnauer hydrogen peroxide sterilizer is undeniable, in order to avoid frequent failures due to operational errors and affect normal use

it is necessary to master the correct operation of the tuttnauer hydrogen peroxide sterilizer Methods.

In order to allow the majority of users to quickly grasp this information, this issue will be mainly introduced in detail.

Hydrogen Peroxide Sterilizer

Correct Operation Specification Of Hydrogen Peroxide Sterilizer

Correct Operation Specification Of Hydrogen Peroxide Sterilizer:

1. Before starting the machine, check the appearance of the tuttnauer hydrogen peroxide sterilizer and whether the power cord is abnormal.

If any abnormality is found, the reason must be checked before starting the machine.

2. Before starting the machine, you must check whether the pump oil of the negative pressure pump is at the normal liquid level, whether the oil color is normal, whether the air filter of the exhaust hole has oil injection, etc.

if the above situation occurs, no matter whether the pump oil is higher or lower The normal liquid level

no matter the oil color is clear and turbid, the pump oil must be replaced, and the oil mist separator must be cleaned or replaced.

3. When water or other volatile substances enter the pump cavity, the ballast valve can be opened for purification

check whether there is water or dirt in the water-gas separator at the inlet, and perform necessary cleaning or replacement.

4. When replacing the pump oil, run the vacuum pump for about three minutes before draining the oil

and then inject about 100 ml of clean pump oil into the oil filling hole for flushing, and then fill the pump oil after the waste oil is drained

generally fill up to the oil Mirror between one-half and two-thirds is fine.

5. If the tuttnauer hydrogen peroxide sterilizer is not used for more than three days, detailed inspection and treatment must be carried out

and the protective plate on the side of the connecting shaft should be removed, and the connecting shaft blade can be rotated by hand.

If the power can’t be moved, the vacuum pump needs to be overhauled.

6. Before disinfection, check whether the door sealing ring of the tuttnauer hydrogen peroxide sterilizer is clean and deformed

check whether the filter screen of the suction and exhaust port of the sterilization chamber is blocked

check whether there is oil and residue on the inner wall; check the plasma emission Whether the head is clean and dry.

Check whether the hydrogen peroxide is too little, not enough for one-time sterilization.